+51 997 777 980Lun–Vie 8:00–18:00 · Lima, Perú
Knauer · ADB74

AZURA® VWD 2.1L Variable Wavelength Detector, fiber optics version

AZURA® VWD 2.1L Variable Wavelength Detector, fiber optics version — Knauer
Marca
Knauer
Código
ADB74
Disponibilidad
Consultar · respuesta en 24 h

without Test and Flow Cell, without Fiber Optic Cables, 190-900 nm Detector Type: Variable single wavelength UV detector, Conductivity monitor, pH Monitor Detection Channels: 4x UV/VIS,1x pH Wavelength Range: 190-900 nm Light Source: High brightness deuterium (D₂) lamp and halogen lamp with integrated GLP chip Flow Cell: Not included, Fiber optic cables are not included Display: Mobile Control (optional) pH Measurement Range: 0-14 Wavelength Accuracy: ± 1.5 nm (verification with integrated holmium oxide filter) Wavelength Precision: 0.4 nm (ASTM E275-93) pH Measurement Accuracy: ± 0.1 (pH 2-12) Wavelength Verification: Internal holmium filter Leak Sensor: Yes, Adjustable sensitivity GLP Report: Lamp recognition, Operating hours, Light source operating hours Maintenance: Frontally accessible deuterium and halogen lamp, as well as flow cell Spectral Bandwidth: 6 nm at Hα line (FWHM) Baseline Noise Level (Test Cell, 254 nm, ASTM E1657-98): ± 11 µAU Drift (254 nm, ASTM E1657-98): 270 µAU/h Linearity (274 nm, ASTM E1657-98): > 2.5 AU Max. Data Rate: 100 Hz (LAN, 1 UV/VIS channel), 5 Hz (LAN, 4 UV/VIS channels with Dl=100 nm) Time Constants: 0.01 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s Integration Time: Automatic

Ficha técnica del fabricante →

Especificaciones técnicas

Detector TypeVariable single wavelength UV detector, Conductivity monitor, pH Monitor
Detection Channels4x UV/VIS,1x pH
Wavelength Range190-900 nm
Light SourceHigh brightness deuterium (D₂) lamp and halogen lamp with integrated GLP chip
Flow CellNot included, Fiber optic cables are not included
DisplayMobile Control (optional)
pH Measurement Range0-14
Wavelength Accuracy± 1.5 nm (verification with integrated holmium oxide filter)
Wavelength Precision0.4 nm (ASTM E275-93)
pH Measurement Accuracy± 0.1 (pH 2-12)
Wavelength VerificationInternal holmium filter
Leak SensorYes, Adjustable sensitivity
GLP ReportLamp recognition, Operating hours, Light source operating hours
MaintenanceFrontally accessible deuterium and halogen lamp, as well as flow cell
Spectral Bandwidth6 nm at Hα line (FWHM)
Baseline Noise Level (Test Cell, 254 nm, ASTM E1657-98)± 11 µAU
Drift (254 nm, ASTM E1657-98)270 µAU/h
Linearity (274 nm, ASTM E1657-98)> 2.5 AU
Max. Data Rate100 Hz (LAN, 1 UV/VIS channel), 5 Hz (LAN, 4 UV/VIS channels with Dl=100 nm)
Time Constants0.01 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s
Integration TimeAutomatic

También en Cromatografia Y Espectrometria

Cotización en menos de 24 h

Cotiza este equipo

Cotizar: AZURA® VWD 2.1L Variable Wavelength Detector, fiber optics version

WhatsApp