Pipette pump capacity250-5000 μL
Pipetting specificationsFor the 5000 μL pipette tip (P/N 417007):
At 250 μL: ± 2.0% accuracy and 3.0% CVAt 500 μL: ±1.5% accuracy and 2.0% CVAt 1000 μL: ±1.0% accuracy and 1.5% CVAt 5000 μL: ±0.5% accuracy and 1.0% CV
Pressure unitPositive pressure unit:Processing: 0-5 Bar (0.1 Bar Increments)Plate dry: 5 Bar at 600 mL/min
Format48-position extraction racks for 3 mL (tabless) columns
24-position extraction racks for 3 and 6 mL (tabless) columns
12-position extraction racks for 10 mL (tabless) columns
12-position extraction racks for 15 mL (tabless) columns
Plate positions6 position upper deck used for sample, extraction plates, pipette tips and solvents.
4 positions on lower deck carousel for multiple elution and waste.
Disposable tipsDisposable sample tips (Position 2): 5000 μL clear tips
Disposable solvent tips (Position 1): 5000 μL clear tips
Disposable solvent tips (Position 6): 5000 μL clear tips
TemperatureOperating temperature: 18°C to 32°C.
Storage and transportation temperature: -25 °C to 60 °C
Humidity10 to 90% RH
Power supply100 to 240 V, 50/60 Hz
FusesT4A at the power inlet (2 required)
Maximum power consumed300 VA
Pressurized air supply>5 L/min (0.18 cubic feet/min), 6 ± 0.2 bar (0.6 ± 0.02 MPa; 87 ± 3 psi)
Inert gas supplyAIR port requirements: >5 L/min (0.18 cubic feet/min), 6 ± 0.2 bar (0.6 ± 0.02 MPa; 87 ± 3 psi).
N2 port requirements: 1 L/min (0.04 cubic feet/min) during processing and 70 L/min (2.5 cubic feet/min) during plate dry, 6 ± 0.2 bar (0.6 ± 0.02 MPa, 87 ± 3 psi)
Vacuum supplyPartial pressure range: < 500 mbar (50 kPa; 7 psi)
Flow rate: ≥ 8.3 L/min (0.29 cubic feet/min)
Max sound level65 dB(A)
GLPGLP software ready
CertificationsCE, CSA certified