LocationA laboratory fume hood with the capacity to handle leakage of solvents and provide suitable airflow to prevent the build-up of flammable vapors.
Ambient temperatureOperating temperature: 15–35 °CStorage and transportation temperature: -25 °C to 60 °C
Humidity20–75% at temperatures up to 31 °C
Electrical supplyEurope: 220–240 V~, 50 Hz (5 A)US and Japan: 100–120 V~, 50/60
Max power consumed1100 VA
Weight42 kg (92.6 lbs)
DimensionsFootprint (W x D x H): 640 x 430 x 640 mm (25.2” x 16.9” x 25.2”)
Max sound level70 dB(A)
PortsEthernet LAN port. Complies with IEEE 802.3 (ANSI 8802-3).USB port. Complies with USB 3.0
InterfacesOn-board computer with 10.4” diagonal touch screen interface.Archiving/back-up is done via USB.Printing is done via LAN.
CertificatesCE, CSA certified
Vacuum pumpMinimum partial pressure: 100 mbar Minimum flow rate: 11 L/min
Temperature increaseTypically 2–5 °C/sec depending on solvent and power applied
Pressure rangeRun at atmospheric pressure
Power range0–120 W at 2.45 GHz
Reactor vial sizes5, 10 and 30 mL
Inert gas (microwave cavity)Approx. 2 L/min (0.07 cubic feet/min); 0.5 bar (0.05 MPa; 7.25 PSI)
Inert gas (manifold option)Approx. 4 L/min (0.14 cubic feet/min); 0.5 bar (0.05 MPa; 7.25 PSI)
Syringe pump3 x digital syringe pumps, with 10 mL sample loop
Reagent bottles5 x 185 mL GL45 glass bottles
Solvent bottles1 x 5 L, 1 x 2 L, and 1 x 1 L, GL45 brown glass bottles
Amino acid rack32 x 30 mL
Waste bottle10L