Power supply100 to 240 V, 50/60 Hz
Maximum power consumed1000 VA
FusesT10A, 250 VAC
TemperatureGas heater temperature range: Ambient to 65 °C, ± 2 °C accuracy.
Plate heater temperature range: Ambient to 80 °C, ± 2 °C accuracy.
Humidity20-90% RH
Gas supply requirementsNitrogen, minimum pressure 3 bar (0.3 MPa, 44 PSI), recommended pressure 3-6 bar (0.3–0.6 MPa, 44–87 PSI), maximum pressure 6.5 bar (0.65 MPa, 94 PSI).
Gas flow: The maximum flow capability for each manifold is 90 L/min. To reach the required flow the pressure must adhere to the recommended pressure.
VentilationFume hood or equivalent enclosure, or connected to a ventilation system via the exhaust outlet. Minimum exhaust capacity m3/min (35.3 ft3/min).
Max sound level75 dB(A)
CertificationsCE marked and CB and NRTL certified
Well plate formatCompatible with well plates that conforms to ANSI SLAS 1-2004 (R2012)